Hanna Instruments New Zealand
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality 360 Value

pH, EC, TDS, Temperature portable meter with CAL Check and Electrode - HI9810-61

pH/EC/TDS/Temperature meter with CAL Check™ and HI1285-61 electrode; EC range: 0 to 6000 µS/cm, TDS range: 0 to 3000 ppm (mg/L)

HI9810-61 is a pH/EC/TDS meter designed to measure pH, μS/cm, mg/L and temperature in hydroponics, greenhouse, farming and groundwater applications. HI9810-61 features Cal Check™, which allows the user to easily check the probe calibration status at any time.

Waterproof

Automatic Temperature Compensation: All readings are compensated for variations in temperature

Low battery indicator

Mediray is Hanna Instruments’ authorised distributor in New Zealand. To purchase our products, please see the details below.

These meters feature a large LCD which displays either pH, EC, TDS or temperature readings along with tutorial instructions. The pH readings are displayed with a 0.1 resolution and an accuracy of +/-0.1 pH while the EC and TDS readings are displayed with a 10 mS/cm and 10 ppm (mg/L) resolution and 2% full-scale accuracy. The temperature correction coefficient (β) is fixed at 2 %/°C and allows for automatic temperature-compensated measurements of EC and TDS. These meters are calibrated manually to a single point with the use of two trimmers. pH is calibrated to pH 7.01 while EC/TDS is calibrated to either 1.41 mS/cm (1413 μS/cm) or 1500 ppm. The LCD screen has a battery life indicator as well
as on-screen tutorial messages.

No probe changes are required when switching your measured parameter between pH, conductivity and TDS. These multiparameter meters reduce the number of instruments required for daily water quality analysis.

The HI9810-61 is supplied with the HI1285-61 pH/EC/TDS/temperature probe with CAL Check. The pH, EC, TDS, and temperature sensor are housed in a single body that connects to the meter with a DIN connector.

Amplified pH electrode: The pH electrode circuit has a built-in amplifier that will reduce the effects of electrical noise on the high-impedance pH measurement. Examples of sources of electrical noise include motors, ballasts, and pumps which are common in greenhouses.

Amperometric EC/TDS sensor: The EC/TDS readings are performed by an amperometric sensor. An alternating voltage is applied to the sensor and the amount of current that passes between the two stainless steel pins is dependent upon the amount of salts (fertilizer) present. A greater amount of salt present results in an increase in conductance.

Polypropylene body: The polypropylene body houses all the sensors in a single-body design and is durable. The probe is gel-filled for maintenance-free operation. It does not have to be refilled periodically.

3 sensors in a single probe

Gel-filled maintenance free pH electrode

Specifications

Specification Name Detail
sku HI9810-61
pH Range 0.0 to 14.0 pH
pH Resolution 0.1 pH
pH Accuracy @ 25°C / 77°F ±0.1 pH
pH Calibration manual, 1-point through offset trimmer
Temperature Range 0.0 to 70.0 °C
Temperature Resolution 0.1°C
Temperature Accuracy @ 25°C / 77°F ±0.5°C
Temperature Calibration Automatic, from 0 to 70°C with ß=2% / °C
EC Range 0 to 6000 µS/cm
EC Resolution 10 µS/cm
EC Accuracy @25°C/77°F ±2% F.S.
EC Calibration manual, 1-point through slope trimmer
TDS Range 0 to 3000 ppm (mg/L)
TDS Resolution 10 ppm (mg/L)
TDS Accuracy @ 25°C / 77°F ±2% F.S.
TDS Conversion Factor 0.5 ppm (mg/L) = 1 µS/cm
TDS Calibration manual, 1-point through slope trimmer
Power Supply 9V / approximately 450 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 145 x 80 x 36 mm (5.7 x 3.1 x 1.4”)
Weight 230 g (8.1 oz.)
EC/TDS Temperature Compensation automatic from 0 to 70°C (32 to 158°F) with β = 2% /°C
Warranty 2 years
   

Ordering info

HI9810-61 is supplied with:

  • HI1285-61 multiparameter probe with CAL Check
  • HI70007 pH 7.01 calibration solution sachet
  • HI70032 1382 ppm (mg/L) calibration solution sachet
  • HI70031 1413 μS/cm calibration solution sachet
  • HI700661 electrode cleaning solution sachets (2)
  • 9v battery (1)
  • Instructions and rugged carrying case

Manuals

Additional information

Weight 0.230 kg

Where to buy

Mediray is Hanna Instruments’ authorised distributor in New Zealand. To purchase our products, please see the details below.

Phone: 0800 633 472

[email protected]

www.mediray.co.nz